Metrohm TitrIC pro III User Manual
Page 14
Application Bulletin 355
Installation Instructions for TitrIC pro III
Page 14 of 14
combination with the MagIC Net method
TitrIC pro
III_titer determination of HCl
.
TitrIC pro III_conductivity calibration:
This method is
used for calibration of the conductivity sensor. It has to
be used in combination with the MagIC Net method
TitrIC pro III_calibration conductivity
.
Import: Click on
Method
at the left (in MagIC Net and
tiamo
TM
), then select under
File / Method manager… / Edit /
Import
… Import the methods related to TitrIC pro III with the
MiPT kit from the MagIC Net CD. You will find the methods
in the folder example/TitrIC of the CD.
Note!
Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo
TM
and MagIC Net
With the TitrIC pro – Combo Kit Partial Loop setup the
following analysis can be carried out using the TitrIC pro
III_MiPT methods in MagIC respectively tiamo
TM
:
The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo
TM
the
method
TitrIC pro III
is selected as a single determination.
Start the determination in autorepeat mode. To set it into
autorepeat go to the Run window, right-mouse-click and
select
Properties Run window
. Under Process select
Autorepeat determination
. Close the window with
. Start
the determination. Now tiamo
TM
is waiting for the start
signal from MagIC Net via RS232.
In the MagIC Net software select the method
TitrIC pro
III_MiPT
. Enter the required sample position, sample
identification etc. into MagIC Net. The Sample volume is
used for the anions injection volume. The cation injection
volume has to be entered into the value 2.
For value 1 choose according to the list below:
Value 1 = 1 complete analysis including titration and
ion chromatography.
Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)
Value 1 = 3 only a titration analysis