3.2 Mounting accessories
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
893 Professional Biodiesel Rancimat
2 Mounting the tubing for the air supply
■
Screw the FEP tubing 250 mm to the air supply connections of
the 893 Professional Biodiesel Rancimat.
3 Inserting the reaction vessel
■
Fill the reaction vessel with the sample.
■
After the required reaction temperature has been reached, insert
the reaction vessel with the mounted reaction vessel cover in the
openings provided on the 893 Professional Biodiesel Rancimat.
4 Establishing the tubing connections
■
Connect the Iso-Versinic
®
tubing that is connected to the measur-
ing vessel cover to the tubing connector of the reaction vessel
cover.
■
Screw the FEP tubing 250 mm which is connected to the M8/olive
tubing adapter of the 893 Professional Biodiesel Rancimat to the
M8/M6 thread adapter of the reaction vessel cover.
3.2.5
Mounting the exhaust air collection tube
The optional exhaust air collection tube (6.2757.000)
[Link target not
found in publication context!]
can be mounted on the 893 Professional
Biodiesel Rancimat for targeted removal of the exhaust air.
NOTE
A total of 8 pieces of Iso-Versinic
®
tubing (6.1839.000) are required in
addition to the exhaust air collection tube.
Proceed as follows to mount the collection tube:
1 Mounting the exhaust air collection tube
■
Insert the exhaust air collection tube with both support pins into
the collection tube holders on the 893 Professional Biodiesel Ran-
cimat in such a way that the connector to the exhaust air removal
is located to the rear.
2 Connecting the measuring vessels
■
Screw the M8/olive tubing adapter into the Out opening of the
measuring vessel cover.
■
Connect one end of the Iso-Versinic
®
tubing to the M8/olive tub-
ing adapter.