beautypg.com

5 technical specifications, 1 general, 2 gold variocell – Metrohm Variocell for Bioscan – 6.5331.1X0 User Manual

Page 18: General, Gold variocell

background image

5.1 General

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

IC Equipment set Variocell for Bioscan

5 Technical specifications

5.1

General

Structure

Flow cell with working, reference and auxiliary electrode.

Cell volume

0.7 µL with the supplied 50 µm spacer.

Working tempera-
ture

The cell should not be operated continuously over 45 °C.

Reference elec-
trode

Type

Solid phase reference electrode

Conversion to
Ag/AgCl

E

Hy-REF

/mV = E

Ag/AgCl

/mV – 328 + 29.9 pH (mV)

5.2

Gold Variocell

Working electrode

Material

Gold

Applications

Sugar and amino acids

Mono-, di-, oligo- and polysaccharides

Sugar alcohols

Sugar amines

Sugar acids

Amino acids

Antibiotics

Working area

Acid medium

-0.35 V…+1.1 V

Alkaline
medium

-1.25 V…+0.75 V