17 connecting the instrument to a computer – Metrohm 940 Professional IC Vario ONE/LPG User Manual
Page 44
3.17 Connecting the instrument to a computer
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
940 Professional IC Vario ONE/LPG (2.940.1150)
■
Remove the two stoppers on the outputs for the two eluent
degassers.
3 Connecting the mixing coil and eluent aspiration tubing in
the low-pressure gradient
■
Screw the long pressure screw for the mixing coil for the low-
pressure gradient (6.2758.020) into the eluent output (Out).
■
Screw a piece of aspiration tubing (6.1834.080) into the output
on each of the two eluent degassers.
■
Screw the mixing valve – eluent degasser connection tubing
(6.1834.120) into eluent input A.
4 Connecting the mixing coil and eluent connection tubing to
the middle plug-in
■
Screw the mixing valve – eluent degasser connection tubing
(6.1834.120) into the output for the eluent degasser.
■
Screw the short pressure screw for the mixing coil for the low-
pressure gradient (6.2758.020) into the high-pressure pump's
inlet valve holder.
Next step
■
Connect each of the two pieces of aspiration tubing (6.1834.080) to
an eluent bottle (see Chapter 3.7, page 22).
3.17
Connecting the instrument to a computer
NOTE
The instrument must be switched off when being connected to a com-
puter.
Accessories
For this step you need the following accessories:
■
USB connecting cable (6.2151.020)