Specifications – Yamato Scientific BM410 Water Baths User Manual
Page 3
1
1. Specifications
Model
BM410
Temp. range during operation
Room temp. +5
℃ ∼ 95℃ *1
Temp. setting range
0
℃∼100℃
Accuracy of temp. adjustment *2
±1℃ at 60℃
Control system
PID Control by Micro Computer
Setting system
Digital Setting By
△▽Key
Display system
Digital display shows the measured/set temp.
Resolution to display/set
1
℃
Operation function
Operation at fixed point
Extra function
Temp. presetting function (preset/select one temp.)
Sensor
Pt100
Ω
Material
SUS316 pipe heater
Heater
Capacity
1kW
Power switch
Also functions as a circuit protector
Other components
Drain (with plug)
Safety device
1. Self-diagnostic Functions
Automatic Overheat Prevention, Sensor Trouble,
TRIAC short, Heater disconnection, Main relay trouble
2. Circuit protector
3. Tank protection cover
4. Temp. fuse
Capacity
7000ml
Tank
Dimension
φ9.8×Depth 5.9 (inches)
φ25×Depth 15 (cm)
External dimensions
(W
×D×H)
*3
12.2
Ч14.2Ч9.1 (inches)
31
Ч36Ч23 (cm)
Weight
15.4 lb. (7kg)
Power source required
AC230V
±10% 50Hz 4.5A
Liquid used
Water
Standard accessories
● Operation manual 1
● Warranty 1
*1 In case of unloaded operation of bath only. The maximum temperature varies on circumstances
and operational conditions.
*2 It shows the performance under rated power supply at the room temperature of 23
℃±5℃ and with
humidity of 65
%±20%.
*3 It does not include projection parts.-
.