Maintenance, Daily maintenance/weekly maintenance, Troubleshooting guide – Yamato Scientific BM410 Water Baths User Manual
Page 18
16
5. Maintenance
Daily maintenance/Weekly maintenance
Do not disjoint the unit
Never modify the unit. There are some high-
voltage areas inside the unit, which could
cause electric shock. Request Yamato
Scientific for internal check, adjustment or
repair.
Maintenance
Daily Maintenance
:
Turn power switch off and disconnect main power cord before attempting any maintenance.
Once the water bath has cooled to room temperature you may empty the water.
Wipe off any excess dirt on the interior and exterior with a damp cloth. Do not use benzene,
thinner or cleanser.
Weekly Maintenance
:
The pipe heater should be cleaned at least twice a month. Accumulation of scale build up
results in poor heat transfer and high sheath temperatures.
When the unit is not in use for a long period of time.
Be sure to disconnect main power and drain the water tank if the unit will not be used for a long
time.
Troubleshooting Guide
Situation
Make Sure
The display on the operation panel
is not lit even when the power
switch is on.
● Check main power cord is connected firmly.
● Check main input voltage.
*When you repeatedly switch on and off, the display does not
light up sometimes. Wait for a few seconds, and switch on
again.
If you have any questions, contact Yamato Scientific.