Sunnex HM2017-44 User Manual
Page 2
Sunnex HM Light Service & Installation Manual
2
1.2 Cleaning and Disinfection
The lamp fixture can be cleaned as needed with a damp cloth and a mild detergent. For disinfection use a
soft cloth with a disinfectant solution such as isopropyl alcohol. NOTE! Do not introduce any fluids to the
inside of the lamp head as this could cause a fire hazard.
The glass lens can be cleaned with a damp cloth and a mild detergent and inside part of the lens can be
easily cleaned with Q-tips.
Technical Data
2.1 Manufacturer
Sales office
Sunnex, Inc.
3 Huron Drive
Natick, MA 01760, USA
Phone: 508-651-0009
Fax: 508-651-0099
Email: [email protected]
Production Facility
Sunnex, Inc.
3 Huron Drive
Natick, MA 01760, USA
Phone: 508-651-0009
Fax: 508-651-0099
Email: [email protected]
2.2 Technical Data
Main Voltage
120VAC 60Hz (230V 50Hz available)
Nominal effect
60W
Safety Class
I
Transformer
Medical Grade Plug-in transformer
INPUT: 120VAC, 60Hz, 0.5A, OUTPUT: +12VAC, 5.0A MAX.
Part# GS-1451
INPUT: 230VAC, 50Hz, OUTPUT: +11.5VAC
Part# 18395 (EURO), 18396 (UK), 18397 (AUSTRALIA)
Bulb
12V/20W/NSP
Color Temperature (CCT)
3584 Kelvin (at 1meter)
Color rendition index (CRI)
95 (at 1meter)
Total irradiance
16 W/m
2
(at 1meter)
Certifications
ETL Listed: CAN/CSA-C22.2 No. 601.1-M90
EN60601-1, UL60601-1, EN60601-2-41
IEC60601-2-41, IEC 60601-1-2 (2001-09) 2nd edition
(EN55011:1991, Group I Class B)
Only original accessories must be used with HM light. The HM light is equipped with a push-button
ON-OFF switch on the lamp head that should be used to turn the light ON and OFF.
EMC
The lamp complies with the EMC requirements in accordance with EN 60601-1-2. Wireless
transmission equipment e.g. mobile phones etc. should not be used in the vicinity of the lamp since
this can affect the function. Special care should be observed during the use of strong sources of
interference e.g. diathermy and the like, so that the diathermy cables not are placed on or close to the
lamp. If in doubt consult your technician or the supplier. Accessories other than here mentioned are
not allowed to use as they may negatively affect emissions and immunity. The complete light must be
recycled in accordance with WEEE standard.