V. troubleshooting, Vi. specifications, Vii. rotators and accessories – SciGene 400 Hybridization Incubator User Manual
Page 12

Model 400 Hybridization Incubator — USER MANUAL
Version 2.1, February 2011 (408) 733-7337, [email protected]
11
V. TROUBLESHOOTING
Symptom
Cause
Solution
Rotator does not turn
after closing door
Rotator control is in the off
position
Turn on rotator.
Door does not shut or is
difficult to open
Latch is improperly
adjusted
Adjust threaded rod
on inside of latch
Temperature is erratic
Thermal controller is
defective
Replace controller
VI. SPECIFICATIONS
Electrical (Oven Unit)
Cat. 1040-60-1 115V AC; 50/60 Hz; 350 W
Cat. 1040-60-2 230V AC; 50/60 Hz; 350 W
Weight
Net
70 lbs (32 kg)
Gross 75 lbs (34 kg) Includes shipping carton
Dimensions (WxDxH)
Interior 14x14x14 inches (36x36x36 cm)
Exterior
18x17x22 inches (45x40x56 cm)
Performance and Controls
Temperature Range
Ambient +5
°C to 99°C
Temperature
Control
± 0.1
°C
Heat
up
Time
3
°C per minute
Temperature Controller
Digital PID, single loop
Temperature Display
Actual/Set dual line LED
Rotation Speed
2 to 20 RPM
Chamber
Stainless Steel
Digital Thermometer Output
Thermocouple
VII. ROTATORS AND ACCESSORIES
Cat. #
Description
UoM
1040-01-0 Hybridization tube, 35x300mm with screw cap
Each
1040-02-0 Hybridization tube, 35x150mm with screw cap
Each
1040-20-5 Replacement Drip Pan for Model 400
Each
1040-21-0 Removable Rocking Platform
Each
1040-21-1 Rotator for 10 large/20 small hybridization tubes
Each
1040-21-2 Maxi-80 Rotator for 80x50ml conical tubes
Each
1040-21-3 Maxi-200 Rotator for 200x15ml tubes
Each
1040-21-4 Maxi-400 Rotator for 400x1.5ml centrifuge tubes
Each
1040-21-5 Maxi-140 Rotator for 40x50mL & 100x15mL plastic tubes.
Each
1040-21-6 Maxi-240 Rotator for 200x1.5mL and 40x50mL tubes.
Each
1040-21-7 Maxi-300 Rotator for 100x15ml tubes & 200x1.5 ml tubes.
Each
1051-52-0 Digital thermometer with NIST certificate of calibration
Each