SciGene 400 Hybridization Incubator User Manual
Page 11

Model 400 Hybridization Incubator — USER MANUAL
Version 2.1, February 2011 (408) 733-7337, [email protected]
10
To remove and replace your Watlow EZ-Zone controller:
a.
Disconnect the power cord from the oven.
b.
Locate the EZ-Zone controller on the front control panel. Push
out the tabs on the left and right side and pull it straight out.
c.
Insert a new controller (SciGene cat. #RP600-0146-07) directly
in its place.
d.
Reconnect the power cord.
e.
Follow the Calibration Check procedure in Section III. F. 1.
f.
Re-calibrate if necessary.
3. Rotator Drive System
The rotator drive system is designed to provide trouble-free
operation for many years and does not require routine
maintenance. However, components of the drive system that fail
due to normal wear and tear over long-term usage can be easily
replaced. Contact SciGene’s Technical Service Department
(
) to obtain replacement parts and perform
these repairs.
B. Cleaning
Clean the exterior and interior surfaces using a mild, detergent-based
spray cleaner and wipe with a soft cloth. Do not use abrasive cleansers
or scouring pads that can scratch the stainless steel.
Under no circumstances should solutions be allowed to enter the
electronics package through the ventilation slots on the top of the unit.