Overview, Specifications – RKI Instruments 65-2513RK User Manual
Page 5

65-2513RK Oxygen Transmitter • 1
Overview
This instruction manual describes the oxygen transmitter. This manual also describes how
to install, start up, configure, maintain, and calibrate the transmitter when using it with a
gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the oxygen transmitter. See the controller operator’s manual for
information specific to the controller.
Specifications
Table 1 lists specifications for the oxygen transmitter.
WARNING: When using the 65-2513RK, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 65-2513RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2513RK as described in this manual.
Table 1:Specifications
Description
Specification
Target Gas
Oxygen (O
2
)
Area Classification
Explosionproof for Class I, Groups B, C, and D
Input Voltage
24 VDC
(24 VDC is typical input. Transmitter will operate from
11 - 30 VDC)
Sampling Method
Diffusion
Signal Output
4 - 20 mA
Detection Range
0 to 25.0% volume
Accuracy
± 0.5% O
2
Response Time
90% in 30 seconds
Operating Temperature
-20° C to 45° C