65-2513RK Oxygen Transmitter • 13

5.

Turn the regulator knob clockwise to close the regulator.

6.

Unscrew the regulator from the calibration cylinder.

7.

Unscrew the calibration cup from the oxygen detector.

NOTE:

For convenience, leave the components of the calibration kit connected by the
sample tubing.

Returning to Normal Operation

1.

Remove the voltmeter leads from the amplifier test points.

2.

Secure the junction box cover to the junction box.

3.

When the display reading rises above the decreasing alarm setpoint, return the
controller to normal operation.

NOTE:

If you do not allow the oxygen reading to increase above the decreasing alarm
point, then unwanted alarms may occur.

4.

Verify that the display reading increases and stabilizes at 20.9% volume.

5.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 6 lists replacement parts and accessories for the oxygen transmitter.

Table 6:Parts List

Part Number

Description

06-1248RK

Sample tubing (3/16 in. x 5/16 in.; specify length when ordering)

18-0405RK-01

Junction box (does not include lid)

18-0406RK

Junction box lid

57-1062RK

Amplifier for oxygen detector

65-2511RK

Oxygen detector, capillary type, conduit-mounting

65-2513RK

Oxygen transmitter (includes detector and amplifier)

71-0111RK

65-2513RK Oxygen Transmitter Operator’s Manual (this document)

81-F301RK-LV

Calibration kit, 34 liter

81-0076RK-01

Zero air calibration cylinder (34 liter)

81-0078RK

Calibration cylinder (100% nitrogen, 17-liter)

81-0078RK-01

Calibration cylinder (100% nitrogen, 34-liter)

81-1050RK

Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter steel
calibration cylinders

81-1117RK

Calibration cup