beautypg.com

RKI Instruments 65-2435RK User Manual

Page 15

background image

65-2435RK CO Transmitter • 15

9. Reconnect the wiring from the controller to the interconnect terminal strip as shown in Table 2

and Figure 3, Wiring the CO Transmitter to a Controller.

10. Reconnect the detector leads to the detector terminal strip as shown in Table 3 and Figure 3,

Wiring the CO Transmitter to a Controller.

11. Reinstall the junction box cover.

12. Turn on or plug in incoming power at the power source end.

13. Turn on the controller.

CAUTION: Allow the sensor to warm up for 5 minutes before you continue with the next step.

14. Calibrate the CO transmitter as described in the Calibration section of this manual.

Replacing the CO Detector

NOTE: In most cases, it is only necessary to replace the CO sensor.

1. Turn off the controller.

2. Turn off or unplug incoming power at the power source end.

3. Remove the junction box cover.

4. Disconnect the detector leads from the detector terminal strip. Note the position of the color-

coded leads as you remove them.

5. Unscrew the detector from the junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the

junction box, then screw the mounting threads of the detector into the conduit hub.

Table 2: Reconnecting the CO Amplifier to a Controller

Amplifier Terminal Strip

Controller Transmitter

Terminal Strip (typical)

FB

4 -20 (FB)

24V

+ V (19 - 30 VDC)

Table 3: Reconnecting the CO Detector to the Amplifier

CO Detector Lead

Amplifier Terminal Strip

Black

BLK

Green

GRN

Brown

BRN

White

WHT

Red

RED