Description, Housing – RKI Instruments 35-3010RKA-07 User Manual
Page 6

2 • 35-3010RKA-07 Sample-Draw Detector
Description
This section describes the components of the
35-3010RKA-07 sample-draw detector
. The
sample-draw detector consists of the housing, flow system, and detection system.
Figure 1: Sample-Draw Detector Component Location
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use. The housing door is
hinged on the left side and is secured by two latches on the right side.
Four mounting feet are attached to the back of the housing (one at each corner). Use the
mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the
bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample-draw
detector’s internal components are mounted to the subpanel.
Hydrophobic
Filter (AcroPak)
Flow Block
for IR CH4
IR CH4
Sensor
H2S Sensor
Am p 1
Pump
Charcoal Filter
LEL T ransm itter
CO Sensor
Com bustible
Sensor
Oxygen Sensor
Am p 2
Inlet F itting For 1/4"
OD Rigid Tubing
Exhaust F itting For 1/4"
OD Rigid Tubing
3/4 Conduit Hub, 2 X
Flow
Block
Main Circuit Board
Oxygen
Transm itter
Reset Switch
Pream p
Circuit
Board