RKI Instruments RX-517 User Manual
Page 12

PT0E-0730
11
<FLOW CHART OF ZERO POINT ADJ>
ZERO MODE DISPLAY
HOLD KEY
ADJ ZERO OK
《Zero set OK》
ZERO SET OK
FAIL ZERO
・Press ZERO key for 3 sec.
Press for 5 sec and move to
the display.
《Zero adj fails》
FRESH AIR?
・When press ENTER key, HC, O2, H2S zero adj
shall start.
・Calibration
ZERO key
ENTER key
ESC key
Move to zero adj
for CO2 adj ESC key
All are OK
HC fails
*Note
・To cancel previous operation, press ESC key.
But can not cancel after press ENTER key.
《RX-517 case》
・For zero adjustment of H2S high, proceed to make zero adjustment by changing to the
display of H2S high
This manual is related to the following products:
See also other documents in the category RKI Instruments Sensors:
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Manual (37 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (45 pages)
- SC-01 (68 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (3 pages)
- GX-94 (52 pages)
- SDM-E2 (89 pages)
- SDM-E2 (224 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer (234 pages)
- Gas Tracer 10,000 ppm (120 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- GX-2009 (2 pages)