RKI Instruments GX-2001 Canadian Version User Manual
Page 29
Model GX-2001 Operator’s Manual
Calibration • 29
Assembling the Calibration Kit
WARNING: Calibrate the Model GX-2001 in a non-hazardous
environment.
1. Attach the adapter plate to the unit. It snaps onto the sensor area.
2. Attach the calibration tubing to the adapter plate, then attach the
opposite end of the tubing to the regulator.
NOTE: Do not attach the regulator to the gas cylinder at this time.
If you are going to calibrate the Model GX-2001 for all the target
gasses at one time, it is recommend that you start with the oxygen
sensor first followed by the H
2
S, CO, and HC sensors in that order.
The reason for this calibration sequence is that while the Model GX-
2001 is in Calibration Mode, you will only be able to toggle through
the target gasses one time, using the MODE/POWER button,
before the unit goes into its start-up sequence.
If you want to go back to a target gas, you will need to turn off the
Model GX-2001 before putting the unit back into Calibration Mode.
HC
O2
Flow
HC
0 2
To Fixed
Flow
Regulator
Calibration Tubing
Adapter Plate
H2S
CO
H2S
C O
RKI
POWER
MODE
DISP
AI R
GX-2001
Figure 6: Assembling the Calibration Kit and Attaching It to the Model GX-2001