General, Cleaning maintenance safety regulations (table), Safety regulations (table) – Retsch AS 200 tap 30.025.0001 User Manual
Page 21: Process action danger

08.01.2008
Retsch GmbH
21
Doc.Nr. D 98.025.9999
General
Cleaning
For thorough, careful, time-saving cleaning of your analytical
sieves we recommend Retsch ultrasonic baths.
Please ask for our special leaflet „Care and Cleaning of Analyti-
cal Sieves“, which is free of charge.
Do not clean the AS 200 tap under running water.
Danger to life through current surge.
Use only a cloth moistened with water.
Solvents are not allowed.
Maintenance
The AS 200 tap is essentially maintenance-free.
Safety regulations (table)
For the AS 200 tap from the various sections
Process
Action
Danger
Safety
Danger to equipment and injury to
persons
arising
from
non-
observance of the safety instruc-
tions
Claims for damages in any form are
excluded.
Packing
Please keep the packing material for
the duration of the guarantee period
A complaint and returning in inade-
quate packing can put your guarantee
claim at risk.
Transport
Do not knock, shake or throw the
AS 200 tap during transport
Electronic and mechanical components
can become damaged.
Temperature variations
In case of temperature variations
protect the AS 200 tap control from
condensed water
Electronic components can become
damaged.
Installation
Install the AS 200 tap only on a
suitable laboratory bench
The machine can produce considerable
unbalance through movement of the
sieved material.
Supplied items
If the delivery is incomplete and/or
there is transport damage you must
inform the transporter and Retsch
GmbH immediately (within 24 hrs)
Later complaints may possibly no
longer be considered.
Ambient temperature
Drops below 5°C
Exceeds 40°C
Electronic and mechanical components
can become damaged.
Performance data can change to an
unknown extent.
Atmospheric humidity
Exceeds 80% at temperatures up to
31°C
Electronic and mechanical components
can become damaged.
Performance data can change to an
unknown extent.
Electrical connection
Power mains system does not agree
with values on type plate
Electronic components can become
damaged.
Cleaning
Remove mains plug before every
cleaning operation.
Danger to life through current
surge
Do not clean under running water
Danger to life through current
surge.