Transport and installation – Retsch AS 200 tap 30.025.0001 User Manual
Page 10

08.01.2008
Retsch GmbH
10
Doc.Nr. D 98.025.9999
Transport and Installation
Packing
Packing is adapted to the transport route and conforms to
the generally applicable packing guidelines.
Please keep the packing material for the duration of the
guarantee period since if there is a complaint and the
machine is returned in inadequate packing your guaran-
tee claim will be at risk.
Transport
After unpacking the AS 200 tap is to be lifted and transported only
by the transport screws TS supplied with the machine. Fig. 1
Net weights:
AS 200 tap approx. 68 kg
AS 200 tap with noise insulating hood
approx. 92 kg
Lifting above head height is not permitted.
The AS 200 tap must not be knocked, shaken or
thrown during transport. Otherwise the electronic and
mechanical components can become damaged.
Temperature variations
If temperature variations are high (e.g. during air
transport) the AS 200 tap must be protected against
condensed water. Otherwise the electronic components
can become damaged.
Intermediate storage
Ensure that the AS 200 tap is also stored dry during in-
termediate storage.
Parameters for the place of installation
Ambient temperature:
5°C to 40°C
If the ambient temperature drops below or exceeds these
values the electrical and mechanical components can become
damaged and performance data are changed to an unknown
extent.
Atmospheric humidity:
Maximum relative humidity = 80% at temperatures up to 31°C, de-
creasing linearly down to 50% relative humidity at 40°C.
Abb. 1
TS