Retsch AS 200 tap 30.025.0001 User Manual
Page 16

08.01.2008
Retsch GmbH
16
Doc.Nr. D 98.025.9999
Starting – Interrupting - Stopping
15 16
Starting:
Press START / button 15 and M simultaneously
Green LED over button 15 lights up
Interrupting (pause function):
Press STOP / button 16 once
Red LED over button 16 lights up
Values remain visible
Press START / button 15 and M simultaneously
Sieving process is continued
Stopping (stand-by function):
Press STOP / button 16 twice
Red LED over button 16 lights up
The complete display is extinguished
Activate displays = press button 15 once
New values can now be eintered
Press START / button 15 a second time
Function as with Starting is carried out
Inserting the sieve stack
The AS 200 tap is suitable for analytical sieves of
200 mm and 203 mm (18“) outside diameter.
In order to loosen the clamping screws easily you
can change the position of the clamping levers
KH independently of the clamping screws. To do
this pull the clamping levers forwards and turn
them both to the required position.
·Then unscrew the two clamping screws C and
push the carriage downwards. Tighten both
clamping screws firmly again.
·Place the sieve stack in the middle of the car-
riage and loosen the two clamping screws.
·Push the carriage upwards until the top edge of
the sieve stack is at the same level as the guide
ring. Then tighten the clamping screws again.
·Lift the tapping arm slightly and place the sieve
lid E with cork plug on the top sieve.
C
E
KH