Retsch DM 200 User Manual
Page 10

16.11.2006
Retsch GmbH
10
Doc.Nr. GB 98.740.9999
The Parameters for the Place of Installation
ambient temperature :
5°C to 40°C
If the temperature is above or below the ambient
temperature, the electronic and mechanical compo-
nent parts may be damaged changing the perform-
ance data in an unpredictable fashion.
humidity :
The maximum relative humidity is 80% at tempera-
tures up to and including 31°C, decreasing linearly to
50% relative humidity at 40°C
If the humidity is higher, the electronic and me-
chanical component parts may be damaged changing
the performance data in an unpredictable fashion.
Installation Altitude:
a maximum of 2,000 m above sea level
Installation
Only place the DM200 on a solid and sturdy founda-
tion.
Electrical Connection
The mains fusing should be designed as follows:
at 3~ 230 and 3/N~ 400V = 16A
at 3~ 115 and 3~ 200V = 20A
you may find the voltage and frequency of the
DM200 on the nameplate.
please ensure that the values agree with the ex-
isting power mains.
only connect the DM200 to the power mains with
the aid of the cable connector supplied.
You may have the connection made directly with a
change-over switch to avoid having to call in an elec-
trician. This has the benefit that the milling disc teeth
may be used by both sides. This is described in the
section "Changing the Direction of Rotation and
Replacing the Milling Disc".
Electrical and mechanical component parts may be
damaged if the values on the nameplate are not
complied with.
!
Modifications or rebuilding it to a supply voltage
other than stated on the machine nameplate should
only be done by an electrician.
Danger of Current Surge!