RTD Temperature Accuracy
The following charts give worst-case temperature accuracy based on stated resistance accuracy of ± (0.015 % + 0.05) Ω.
Pt100 TCR=0.00385
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Pt100 TCR=0.003902
-200 to 648.9 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
E
rro
r °
C
Pt100 TCR=0.003916
-200 to 648.9 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
E
rro
r °
C
Pt100 TCR=0.003926
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Cu10 TCR=0.00427
-200 to 260 °C
1.2
1.25
1.3
1.35
1.4
-200
-100
0
100
200
300
Temperature °C
Erro
r °C
Cu50 TCR=0.00428
-50 to 150 °C
0
0.1
0.2
0.3
0.4
0.5
-50
0
50
100
150
Temperature °C
Er
ro
r °
C
Ni120 TCR=0.00672
-80 to 260 °C
0.07
0.08
0.09
0.1
-100
-50
0
50
100
150
200
250
300
Temperature °C
Error °C
Ni110 TCR=0.00530
-100 to 260 °C
0.1
Error °C
0.08
0.1
0.12
4
0.16
-100
0
100
200
300
400
Temperature °C