T/c temperature accuracy, Model 525 operating instructions – Practical Instrument Electronics 525 Rev B User Manual
Page 7

Model 525 Operating Instructions
82 E. Main Street Suite 3.14 · Webster, NY 14580
Tel: 585-872-9350 · Fax: 585-872-2638
[email protected] · http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright ã 2011. All rights reserved.
525-9002 7/13/11
7-8
T/C Temperature Accuracy
The following charts give worst-case temperature accuracy based on stated millivolt accuracy of ± (0.008 % + 0.006). Temperature
is uncompensated on the horizontal axis, referenced to 0 °C. Cold Junction calibration accuracy of 0.1 °C is not included in the
temperature error.
TYPE B
594 to 1820 °C
0
0.2
0.4
0.6
0.8
1
1.2
500
700
900
1100
1300
1500
1700
1900
Temperature °C
Er
ro
r °
C
TYPE E
-260 to 1000 °C
0
0.2
0.4
0.6
0.8
1
1.2
1.4
-400
-200
0
200
400
600
800
1000
1200
Temperature °C
Er
ro
r °
C
TYPE N
-229 to 1300 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Er
ro
r °
C
TYPE R
24 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Er
ro
r °
C
TYPE S
21 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Er
ro
r °
C
TYPE T
-251 to 400 °C
0
0.2
0.4
0.6
0.8
1
1.2
-300
-200
-100
0
100
200
300
400
500
Temperature °C
Er
ro
r °
C
TYPE K
-245 to 1372 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
1600
Temperature °C
Er
ro
r °
C
TYPE J
-210 to 1200 °C
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Er
ro
r °
C