User interface changes – New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual
Page 3

NE-1000X
3
www.SyringePump.com
These functions will operate normally, but when the next RET function is executed after the Program Branch function, program
operation will return to the Program Phase immediately following the Phase where the Pumping Program branched, or the Program
Phase that was interrupted for events:
IF:nn, JP:nn
After one of these functions cause a Program Phase Branch to Phase 'nn', a RET function will return the Program to the Phase
following the function that caused the branch, thus converting these Pumping Program branch functions into program sub-routine call
functions.
ET:nn, ES:nn
These functions create a background Event trap. When the Event trap is triggered, the Pumping Program immediately jumps to
Program Phase 'nn'. A RET function will return to the Program Phase that was interrupted by the Event trap. NOTE: When the
Pumping Program returns to the interrupted Phase, the Program Phase starts over and not at the point that was interrupted. For
example, if the interrupted Phase specified a pumping volume, on return, the entire pumping volume will be dispensed, not just the
remainder of volume from the point of interruption.
RS-232Remote Support:
Return function command:
FUN RET
===============================================================================================
2.
User Interface Changes
===============================================================================================
Re-ordering of Program Function Selection Menu
The Program Function Selection menu has been broken up into a top menu and a control function sub-menu for easier entry.
The top menu is as follows:
BASE
Base Rate and Dispense Volume Function
RATE
Rate and Dispense Volume Function
FILL
Fill Function
RECP
Reciprocating pumping Function
INCR
Rate increment and Dispense Time Function
DECR
Rate decrement and Dispense Time Function
LINR
Linear Ramp Rate and Dispense Time Function (1st and 2nd halves)
PS:00
Pumping Pause Timer Function
STOP
Program Stop Function
CTRL
Select Control Functions Sub-menu
Control Function sub-menu is as follows:
nn
JP:01
Jump to phase number
LP:ST
Loop start
LP:EN
Continuous loop end
nn
LP:01
Loop for a set number of iterations
PR:IN
Query user for sub-program number
nn
PR:01
Define start of sub-program
nn
IF:01
If programmable input logic pin is low, jump to Phase number
nn
ET:01
Set background event trap to jump to Phase number
nn
ES:01
Set background event square input event trap to jump to Phase number
nn
SE:01
Set Stall Event background event trap to jump to Phase number