3.1 Initial start-up

When the bath is started up for the first time, it should be supervised continuously until steady conditions
have been reached.

3.2 Installation - waterbaths

Mains
connection

The mains connection cable must be placed away from any hot surface.

Filling

In order to garant the high quality of the stainless steel it is recommended by the

steel producer to fill the waterbath with deionised water.

Draining

The tank can be drained through the drain valve. The thermostating fluid must be cooled
to room temperature before draining. Drain the fluid either directly into an appropriate
container or connect an appropriate hose to the drain valve.

Operation with non-flammable thermostating liquids only!

drain valve

IMPORTANT! The water level in the bath must be between the

two markings at the right side of the tub!

Waterbaths can be fitted with a

level control system (see section 9).

NOTE: Polluted fluids may not be drained directly into the public

wastewater system. If the thermostating fluid is polluted, please

make sure that it will be properly filtered or dumped according to the

applicable regulations.

max water level

min water level