Memmert ONE 7 - 45 Oilbath (Generation 2003) User Manual
Page 13

EXCELLENT
page 13
4.3.3 Hold time of the setpoint temperature
The bath switches off the heating after
the end of the selected hold time. In this
case the hold time includes the heating-
up time..
Adjustment range:
1
min bis 999 hrs
Setting accuracy:
<100 hrs: 1 min
≥100 hrs: 1 hr
Indication accuracy:
<10 hrs: 1 min
≥10 hrs: 1 hr
Rotate the push/turn control until the (hold) and the symbol flashes.
The duration of the hold time can then be selected with the SET key depressed,
as described in Section 4.3. After the SET key has been released the bath briefly
flashes the hold time setting.
If no hold time has been programmed, the hold time is activated immediately
and the display alternates between the actual temperature and the running time
of the hold time. As with switch-on delay, the residual time is shown with a
negative sign and runs down.
If no hold time is required, it can be de-activated by the setting OFF.
°C
°C
°C
4.3.4 Setpoint-dependent waiting time
The bath switches off the heating after the selected
hold time has elapsed. The hold time begins only
when the setpoint temperature has been reached.
Setting:
ON
or OFF
Rotate the push/turn control until
flashes.
With the SET key depressed, the setpoint-dependent waiting time can be activated
or de-activated.
°C
°C
*
t
90°C
hold 6h
*
t
hold 6h
90°C