samples containing phenol or phenol chloroform, the rotor
should be cleaned immediately after use.

5.4 Disinfection

Should a spill of infectious materials occur within the rotor or
chamber, the unit should be disinfected. This should be per-
formed by qualified personnel with proper protective equipment.

5.5 Replacing fuses

Check the fuse when it is recommended in the Troubleshooting
Guide located in this manual. The fuse holder is located in the
power inlet on the rear of the unit. Disconnect the power cord
from the power inlet. Open the fuse holder drawer by inserting a
small screwdriver under the tab and prying it open. Remove the
innermost (operative) fuse from its retaining tabs and replace the
fuse if necessary. A spare fuse is located in the outermost
chamber of the fuse drawer. Replace only with a fuse of exactly
the same value as the original. (Fuse type may be found in the
Technical data section of this manual.)

6. Troubleshooting Guide

Please refer to this guide before calling for service.

Centrifuge will not start

Possible reason:

No power supply

Solution:

Check that power is being supplied to
the outlet
Check that the power cord is plugged
into both the wall outlet and the back of
the centrifuge
Check that power cord is not damaged

Possible reason:

Blown fuse

Solution:

Check fuse and replace if necessary

Lid lock will not release

Possible reason:

Defective lid lock

Solution:

Open manually and have unit serviced

Possible reason:

No power from PC board

Solution:

Call for service

Possible reason:

Lid lock is jammed

Solution:

Call for service

9