TROUBLESHOOTING GUIDE (continued)

PROBLEM

POSSIBLE SOLUTION

Temperature is too low

If door has opened, unit may not have recovered yet.
Confirm temperature set point.

Unit will not heat above temperature

Confirm set point. Check temperature of chamber

that is below set point

with a thermometer and recalibrate if needed.

Temperature display and reference

Be sure that unit has been allowed to stabilize for

thermometer do not match

1 hour. Thermometer should be suspended in
rotating bottle and not touching bottle surface. Only
certified reference thermometers should be used.

Can not adjust set point or calibration

Turn unit off for 5 seconds to reset. If problem
persists call service

Unit calibrated at one temperature

This can be a normal condition if temperatures or

load not at another

vary widely. For best accuracy, calibrate at set point

TECHNICAL SUPPORT / SERVICE

Should your ProBlot hybridization oven require service, call Labnet International’s Service Depart-
ment at 732-417-0700 between the hours of 9:00am and 5:00pm EST. If service necessitates the
return of the unit to Labnet’s Service Department, a return authorization (RA) number must first be
obtained and a decontamination certificate completed. The unit must be thoroughly decontami-
nated before shipping. The unit must then be sufficiently packed for shipment. Any shipping dam-
age on returned units will be the customer’s responsibility.

ACCESSORIES

B3003

Large 300x35mm bottle w/cap

B1500-CAP

Cap for 35mm bottles

B3003-PC

as above with safety coating

B2500-SEAL Seal for all 35mm bottles

B3003-DC

Large bottle with caps on both ends B1124

Rack for 6 35mm bottles

B2503

Medium 250x35mm bottle

H9088

Mesh 23x23cm (pack of 5)

B1503

Small 150x35mm bottle w/cap

H9089

Mesh 10x15cm (pack of 5)

B1503-PC

as above with safety coating

B0753

Small 75x35mm bottle

B2753

Small bottle with 2 x75mm chambers B1507

150x70mm bottle, vented

H1200-RA

Rocking platform

B3007

300x70mm bottle, vented

H1200-250A Clamp platform 6 x 250ml flasks

H1204-80A

Rotisserie 4 x 70mm bottles

H1212-40VA Rotisserie 12 x 50ml tubes, vertical