TROUBLESHOOTING GUIDE (continued)
PROBLEM
POSSIBLE SOLUTION
Temperature is too low
If door has opened, unit may not have recovered yet.
Confirm temperature set point.
Unit will not heat above temperature
Confirm set point. Check temperature of chamber
that is below set point
with a thermometer and recalibrate if needed.
Temperature display and reference
Be sure that unit has been allowed to stabilize for
thermometer do not match
1 hour. Thermometer should be suspended in
rotating bottle and not touching bottle surface. Only
certified reference thermometers should be used.
Can not adjust set point or calibration
Turn unit off for 5 seconds to reset. If problem
persists call service
Unit calibrated at one temperature
This can be a normal condition if temperatures or
load not at another
vary widely. For best accuracy, calibrate at set point
TECHNICAL SUPPORT / SERVICE
Should your ProBlot hybridization oven require service, call Labnet International’s Service Depart-
ment at 732-417-0700 between the hours of 9:00am and 5:00pm EST. If service necessitates the
return of the unit to Labnet’s Service Department, a return authorization (RA) number must first be
obtained and a decontamination certificate completed. The unit must be thoroughly decontami-
nated before shipping. The unit must then be sufficiently packed for shipment. Any shipping dam-
age on returned units will be the customer’s responsibility.
ACCESSORIES
B3003
Large 300x35mm bottle w/cap
B1500-CAP
Cap for 35mm bottles
B3003-PC
as above with safety coating
B2500-SEAL Seal for all 35mm bottles
B3003-DC
Large bottle with caps on both ends B1124
Rack for 6 35mm bottles
B2503
Medium 250x35mm bottle
H9088
Mesh 23x23cm (pack of 5)
B1503
Small 150x35mm bottle w/cap
H9089
Mesh 10x15cm (pack of 5)
B1503-PC
as above with safety coating
B0753
Small 75x35mm bottle
B2753
Small bottle with 2 x75mm chambers B1507
150x70mm bottle, vented
H1200-RA
Rocking platform
B3007
300x70mm bottle, vented
H1200-250A Clamp platform 6 x 250ml flasks
H1204-80A
Rotisserie 4 x 70mm bottles
H1212-40VA Rotisserie 12 x 50ml tubes, vertical