To operate the rocking platform, turn the rotisserie power switch to the ON position and
adjust the speed using the rotisserie speed knob.
CARE & MAINTENANCE
No routine maintenance is requires for the electrical or mechanical components of the unit.
The incubator exterior, interior, rotisserie and shaking or rocking platform (if present)
should be wiped down periodically with a soft damp cloth with mild soap. Do not use chlo-
rine-based bleach or abrasives. Any spills in the incubator and/or on the rotisserie, shaking
or rocking platforms should be cleaned up immediately (see Decontamination). Be sure to
disconnect the power cord before cleaning or decontaminating the oven.
DECONTAMINATION
Should a spill occur in the oven, it is easily cleaned by wiping the affected area first with a
dilute detergent solution of IsoClean or CountOff. The area should then be cleaned with
distilled water. This method works on the oven interior, drip tray, shaker or rocker platform
and the rotisserie. Bottles and caps may also be soaked in a dilute decontaminating deter-
gent such as IsoClean or CountOff and then rinsed in distilled water.
TROUBLESHOOTING GUIDE
PROBLEM
POSSIBLE SOLUTION
Oven will not power up or not heat
Check power cord, outlet and unit circuit breaker
Rotisserie will not turn
Check rotisserie On/Off switch and speed setting
Rotisserie speed erratic, jumps
Check for balanced load on rotisserie. Even num-
ber of bottles loaded in opposing positions
Shaking platform will not operate
Check timer set to RUN or a time setting
Check that nothing is blocking shaker platform
sides
Temperature too high
Check set point and readjust if necessary
Check calibration
Chamber temperature goes above set
Normal operation in initial heat up or if door
point and settles back to set point
opened for a long period
Temperature will not remain stable
Check that set point is at least 5C above ambient
or display shows “LO”
which is minimum set and operating point
Indicated temperature is unstable
A slight variation of +/-0.1C is normal. Larger
fluctuations may be ambient variations from drafts,
door opening and closing, a fan obstruction or
failure or electrical noise from RFI (motors, etc.)