beautypg.com

Labnet Deluxe ProBlot Hybridization Systems (6, 12, 12S and 24) User Manual

Page 7

background image

To operate the rocking platform, turn the rotisserie power switch to the ON position and
adjust the speed using the rotisserie speed knob.

CARE & MAINTENANCE
No routine maintenance is requires for the electrical or mechanical components of the unit.
The incubator exterior, interior, rotisserie and shaking or rocking platform (if present)
should be wiped down periodically with a soft damp cloth with mild soap. Do not use chlo-
rine-based bleach or abrasives. Any spills in the incubator and/or on the rotisserie, shaking
or rocking platforms should be cleaned up immediately (see Decontamination). Be sure to
disconnect the power cord before cleaning or decontaminating the oven.

DECONTAMINATION
Should a spill occur in the oven, it is easily cleaned by wiping the affected area first with a
dilute detergent solution of IsoClean or CountOff. The area should then be cleaned with
distilled water. This method works on the oven interior, drip tray, shaker or rocker platform
and the rotisserie. Bottles and caps may also be soaked in a dilute decontaminating deter-
gent such as IsoClean or CountOff and then rinsed in distilled water.

TROUBLESHOOTING GUIDE

PROBLEM

POSSIBLE SOLUTION

Oven will not power up or not heat

Check power cord, outlet and unit circuit breaker

Rotisserie will not turn

Check rotisserie On/Off switch and speed setting

Rotisserie speed erratic, jumps

Check for balanced load on rotisserie. Even num-
ber of bottles loaded in opposing positions

Shaking platform will not operate

Check timer set to RUN or a time setting

Check that nothing is blocking shaker platform
sides

Temperature too high

Check set point and readjust if necessary
Check calibration

Chamber temperature goes above set

Normal operation in initial heat up or if door

point and settles back to set point

opened for a long period

Temperature will not remain stable

Check that set point is at least 5C above ambient

or display shows “LO”

which is minimum set and operating point

Indicated temperature is unstable

A slight variation of +/-0.1C is normal. Larger
fluctuations may be ambient variations from drafts,
door opening and closing, a fan obstruction or
failure or electrical noise from RFI (motors, etc.)