Guidance and manufacturer’s declaration – electromagnetic immunity
The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the
NCT1 should assure that it is used in such an environment.
Immunity test
Compliance level Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
6 kV contact
8 kV air
6 kV contact
8 kV air
Not applicable
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
Not applicable
IEC 60601
test level
2 kV for power
supply lines
1 kV for input/
output lines
1 kV line(s) to
line(s)
2 kV line(s) to
earth