Emc table – Kinetik Thermometer NCT1 User Manual
Page 19

18
EMC Table
Guidance and manufacturer’s declaration – electromagnetic immunity
The NCT1 is intended for use in the electromagnetic environment specified below. The customer or the user of the NCT1
should assure that it is used in such an environment.
Immunity test
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Not
applicable
3 V/m
IEC 60601
test level
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz
to 2,5 GHz
Portable and mobile RF communications equipment should be used no closer
to any part of the NCT1, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey, a should be less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following
symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 1,2
d = 1,2
d = 2,3