Emc table – Kinetik Thermometer NCT1 User Manual
Page 20

19
EMC Table
Guidance and manufacturer’s declaration – electromagnetic immunity
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the NCT1 is used exceeds
the applicable RF compliance level above, the NCT1 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the NCT1.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.