Kds35 small animal ventilator, Pump operation – KD Scientific 35 Ventilator User Manual

Page 9

background image

7

KDS35 Small Animal Ventilator

Pump Operation:

CONNECTION TO THE ANIMAL

The vertical slide valve has four ports (see figure below):

•The top port (furthest from the chassis in new orientation) is the return channel from

the animal, which leaves the Pump at the port on the side. It is at this top port that

expired air can be collected or recycled.

•The middle port is the room air/gas channel to the animal.

•The bottom port (closest to the chassis in new orientation) is the room air or gas

mixture intake.

NOTE: It is important that the tubes to and from the animal be as short as possible. The new

orientation allows the animal to be as close as possible to the Ventilator for minimal dead

space in the system.

The actual connection to the animal can be either:

•Tracheal cannula with y adapter

•Intubation cannula with y adapter

A selection of trachael cannula and endotracheal tubes are listed in the KD Scientific

Catalog and on our website www.kdscientific.com. It is important that the volume of tubing

shared by inspired and expired air (dead space) be kept to a minimum by having the “Y” or

“T” connector as close to the animal as possible.