beautypg.com

Kds35 small animal ventilator, Pump operation – KD Scientific 35 Ventilator User Manual

Page 11

background image

9

KDS35 Small Animal Ventilator

Pump Operation:

(continued)

SETTING VOLUME

The volume delivered is read by observing the excursion of the O-ring, on the piston, which is

closest to the valve block. This piston O-ring excursion is relative to the volume scale on the

outside of the cylinder wall. Volume change is accomplished while the Ventilator is running by

turning “volume” knob in the desired direction.

SETTING RATE

Adjust the “Rate” control until the desired rate appears on the LED display. Avoid running the

Ventilator at rates exceeding 150 breaths/min for periods in excess of 2 hours.

CAUTION - DO NOT OVER INFLATE THE ANIMAL. A GOOD RULE OF THUMB IS TO

OBSERVE THE DEPTH OF RESPIRATION IN THE NORMAL CALM UNANESTHETIZED

ANIMAL AND ADJUST THE VOLUME TO APPROXIMATE THIS RESPIRATION DEPTH

WHEN THE ANIMAL IS ATTACHED TO THE VENTILATOR.

USE OF GAS MIXTURE

It is possible to introduce non-explosive air/gas mixtures into the lower room air port.

However, the source of gas mixture is usually from a tank delivering gas is at a constant rate

while the Ventilator takes in air/gas intermittently. Some sort of buffer/expansion chamber

must be provided between the tank and the Ventilator. KD Scientific offers a connection kit

(73-0397) to depressurize the gas prior to entering the ventilator.

This Ventilator is not

designed to take air/gas inputs under pressure.

CREATING A POSITIVE END-EXPIRATORY PRESSURE (PEEP)

To create a PEEP pressure when using the ventilator, simply connect a length of tubing to the

“Exhaust” port and submerge the other end in a column of water. There are two locations in

which the clip and cylinder can be installed - depending on the chosen orientation. The depth

of the end of the tubing (in cm) below the water surface will give the PEEP pressure in

cmH2O.