Kds model 100 series, Specifications – KD Scientific 100 Series User Manual
Page 4

2
KDS Model 100 Series
Specifications
100 Series, Including KDS100, 101 and 120
Syringe Size
10 microliter to 60 milliliter
Electrical Rating
US model 115V~, 0.1A
CE model 230V~, 0.06A
Fuse
5x20 mm, 250V~ slow blow, 0.1A
Voltage Operating Range
US model 100-120V~, 50/60Hz
CE model 200-240V~,50/60Hz
Drive Mechanism
microprocessor controlled stepper gearmotor
(
1
⁄
2
microcrostepping), driving a leadscrew through
a belt and pulley drive mechanism
Pusher Advance per Half Step
~0.529 micron or 2.083x10
-5
inch (KDS100)
~0.088 micron (KDS101)
Linear Travel
KDS100, KDS120
minimum 0.00635 cm/hr; maximum 76.18 cm/hr
KDS101
minimum 0.00106 cm/hr; maximum 12.2 cm/hr
Minimum Stepping Rate
one
1
⁄
2
step in thirty seconds – may vary depending
on syringe size
Maximum Stepping Rate
400
1
⁄
2
steps / second
Speed Range
12000:1
Flowrate Range
0.1µl/hr (10 µl syringe) to 506ml/hr
(60 ml syringe)
Linear Force
20 lb min. ( 9 Kgm)
Dimensions
9x6x5 inch (23x15.25x13 cm)
Weight
4.5 lb (2 Kgm)
Atmospheric Specifications
Temperature
4˚C to 40˚C (40˚F to 104˚F)
Humidity
20% to 80% RH, non-condensing
Mode of Operation
Continuous
Classification
Class I
Pollution Degree
2
Installation Category
II
Output
N/A
Physiological Effects
N/A
Cooling Conditions
No special considerations
Mechanical Stability
No special considerations
Protective Packaging
No special considerations
Earth Terminals
No External connections required
Removable Protective Means
N/A
Supplier Name
KD Scientific Inc.
Address
84 October Hill Road, Holliston, MA 01746