Kds model 100 series – KD Scientific 100 Series User Manual
Page 12
![background image](/manuals/564061/12/background.png)
10
KDS Model 100 Series
KDS120 PUSH-PULL PUMP
Travel Direction
Withdraw Clamp
Release Button
Withdraw Bracket Clamp
Syringe Rest
Syringe Clamp
To facilitate loading , the pusher block can be released from the leadscrew by pressing the
bronze release button (1) and can be manually moved along the guide rods to accommo-
date the syringes.
First place the withdraw syringe in the withdraw syringe holder, using the spring loaded
retaining clamp to retain the syringe barrel in the V of the syringe holder.
Make sure the barrel flange is held firmly by the withdraw clamp and that the adjusting
screws on the withdraw clamp are firmly tightened.
While keeping the button (1) pressed firmly "in" slide the block along the guide rods so that
the syringe plunger flange can be retained by the clamp on the pusher block. Tighten the
locking screws on the adjusted syringe clamp.
Loosen the locking screw on the Adjustable syringe rest and position the rest so that the
plunger of the loaded syringe is pressed firmly against the pusher block and the syringe
barrel flange is pressed up to the syringe rest.
The barrel of the loaded, infusion syringe should be clamped in the V of the syringe holder
with the spring loaded clamp.
Tighten the screw on the syringe rest to lock the rest in place on the guide rod.