KDS Model 100 Series

14

MODELS KDS100 AND KDS120

Syringe size

Minimum

Maximum

10µl

0.1µl/h

126µl/h

25µl

0.1µl/h

318µl/h

50µl

0.2µl/h

625µl/h

100µl

1.0µl/h

1274µl/h

250µl

2.0µl/h

3164µl/h

500µl

3.0µl/h

6359µl/h

1ml

0.01ml/h

13.2ml/h

2.5ml

0.02ml/h

31.7ml/h

3ml

0.02ml/h

44.9ml/h

5ml

0.03ml/h

87.0ml/h

10µl

0.1µl/h

126µl/h

10ml

0.1ml/h

125ml/h

20ml

0.1ml/h

219ml/h

30ml

0.1ml/h

282ml/h

60ml

0.2ml/h

426ml/h

MODEL 101

Syringe size

Minimum

Maximum

10µl

0.001µl/m

0.382µl/m

25µl

0.001µl/m

1.010µl/m

50µl

0.001µl/m

1.762µl/m

100µl

0.001 µl/m

3.542 µl/m

250µl

0.01 µl/m

8.78 µl/m

500µl

0.01µl/m

17.65µl/m

1ml

0.1 µl/m

35.2 µl/m

2.5ml

0.1 µl/m

88.0 µl/m

3ml

0.1 µl/m

122.5 µl/m

5ml

0.1 µl/m

176.2 µl/m

10ml

0.001 ml/m

0.351 ml/m

20ml

0.001 ml/m

0.602 ml/m

30ml

0.001 ml/m

0.773 ml/m

60ml

0.001 ml/m

1.175 ml/m

Note: Syringes from different manufacturers can have slightly different limits.