Kds model 100 series, Standard minimum and maximum flow rates – KD Scientific 100 Series User Manual
Page 16
![background image](/manuals/564061/16/background.png)
KDS Model 100 Series
14
MODELS KDS100 AND KDS120
Syringe size
Minimum
Maximum
10µl
0.1µl/h
126µl/h
25µl
0.1µl/h
318µl/h
50µl
0.2µl/h
625µl/h
100µl
1.0µl/h
1274µl/h
250µl
2.0µl/h
3164µl/h
500µl
3.0µl/h
6359µl/h
1ml
0.01ml/h
13.2ml/h
2.5ml
0.02ml/h
31.7ml/h
3ml
0.02ml/h
44.9ml/h
5ml
0.03ml/h
87.0ml/h
10µl
0.1µl/h
126µl/h
10ml
0.1ml/h
125ml/h
20ml
0.1ml/h
219ml/h
30ml
0.1ml/h
282ml/h
60ml
0.2ml/h
426ml/h
MODEL 101
Syringe size
Minimum
Maximum
10µl
0.001µl/m
0.382µl/m
25µl
0.001µl/m
1.010µl/m
50µl
0.001µl/m
1.762µl/m
100µl
0.001 µl/m
3.542 µl/m
250µl
0.01 µl/m
8.78 µl/m
500µl
0.01µl/m
17.65µl/m
1ml
0.1 µl/m
35.2 µl/m
2.5ml
0.1 µl/m
88.0 µl/m
3ml
0.1 µl/m
122.5 µl/m
5ml
0.1 µl/m
176.2 µl/m
10ml
0.001 ml/m
0.351 ml/m
20ml
0.001 ml/m
0.602 ml/m
30ml
0.001 ml/m
0.773 ml/m
60ml
0.001 ml/m
1.175 ml/m
Note: Syringes from different manufacturers can have slightly different limits.
Standard Minimum and Maximum Flow Rates