Kds model 100 series, Operating instructions – KD Scientific 100 Series User Manual
Page 10

8
KDS Model 100 Series
Operating Instructions
(Continued)
CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING
While the pump is running it is possible to look at the volume and flowrate settings without
interrupting the operation of the pump. If required, the volume and flowrate settings can be
changed while the pump continues to run. When the changes are entered the pump immedi-
ately changes to the new settings.
1)
While operating, press select to return to the main menu.
2)
Select Volume to display the set dispense volume. If no change is required
press select.
3)
For a volume change use the arrow keys and enter with the select key.
4)
The display moves to RATE, permitting a change if required.
Press select if no change is required, or use the arrow keys to change the
setting. The select key immmediately changes to the new flowrate and the
volume display continues to increment, uninterrupted to the new
dispense setting.
Note: If the VOLUME is changed to a setting smaller than the volume already
accumulated then the pump stops when the new, smaller dispense volume is
selected.
CHANGE OR REVIEW RATE SETTING WHILE RUNNING
1)
Press select to change the volume accumulator display to the main menu.
2)
Select RATE.
3)
Make rate changes, if required, and press select.
The pump immediately changes to the new flow rate and displays the continuing vol-
ume accumulation uninterrupted by the change in flow rate.
FAST FORWARD
Press the run and the
keys simultaneously to actuate the fast forward mode. The pump
travels at its maximum rate while both keys are pressed simultaneously.
➔