KDS Model 100 Series

Operating Instructions

(Continued)

CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING

While the pump is running it is possible to look at the volume and flowrate settings without

interrupting the operation of the pump. If required, the volume and flowrate settings can be

changed while the pump continues to run. When the changes are entered the pump immedi-

ately changes to the new settings.

1)

While operating, press select to return to the main menu.

2)

Select Volume to display the set dispense volume. If no change is required

press select.

3)

For a volume change use the arrow keys and enter with the select key.

4)

The display moves to RATE, permitting a change if required.

Press select if no change is required, or use the arrow keys to change the

setting. The select key immmediately changes to the new flowrate and the

volume display continues to increment, uninterrupted to the new

dispense setting.

Note: If the VOLUME is changed to a setting smaller than the volume already

accumulated then the pump stops when the new, smaller dispense volume is

selected.

CHANGE OR REVIEW RATE SETTING WHILE RUNNING

1)

Press select to change the volume accumulator display to the main menu.

2)

Select RATE.

3)

Make rate changes, if required, and press select.

The pump immediately changes to the new flow rate and displays the continuing vol-

ume accumulation uninterrupted by the change in flow rate.

FAST FORWARD

Press the run and the

keys simultaneously to actuate the fast forward mode. The pump

travels at its maximum rate while both keys are pressed simultaneously.