JULABO MB Heating Immersion Circulator User Manual
Page 44
Remote control
44
12.3. List of commands
out commands: Setting parameters or temperature values.
Command
Parameter
Response of circulator
version None
Number
of
software version (V X.xx)
status none
Status
message, error message (see page 45 )
out_mode_01 0
Use
working
temperature >t 1<
out_mode_01 1
Use
working
temperature >t 2<
out_mode_01 2
Use
working
temperature >t 3<
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_sp_00 xxx.xx Set
working temperature. „t 1“
out_sp_01 xxx.xx Set
working temperature. „t 2“
out_sp_02 xxx.xx Set
working
temperature. „t 3“
out_sp_03
xxx.xx
Set high temperature warning limit
„t High“
out_sp_04
xxx.xx
Set low temperature warning limit
„t Low“
out_par_06 xxx
Xp
control
parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.
in commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04 none Setpoint
temperature („SafeTemp“) of the excess
temperature protection
in_sp_00 none Working
temperature „t 1“
in_sp_01 none Working
temperature „t 2“
in_sp_02 none Working
temperature „t 3“
in_sp_03 none High
temperature warning limit „t High“