JULABO MB Heating Immersion Circulator User Manual
Page 36

Menu functions
36
 
9.4.1. ATC STATUS - YES / NO 
In the second submenu the ATC function for the temperature sensor 
selected above is activated >YES< or deactivated >NO<. 
>YES< (factory setting) The controller of the circulator uses the
original curve of the temperature sensor or the new curve 
measured during the ATC calibration. 
Important: Set to >NO< during the calibration process 
 
>NO< An ATC calibration is to be carried out.
Important: Set to >YES< after calibration.
In the > ATC STATUS < >YES< the ATC calibration always affects 
the current working temperature; also the one set via interface. 
 
 
 
 
9.4.2. ATC - TYPE: 1 -/ 2 -/ 3 POINT 
A >1-point<, >2-point< or >3-point< calibration can be carried out.
First geometrically define the location for calibration (measuring point 
CT), then determine the temperature values of the calibration points. 
The type of calibrations also determines the number of the following 
pairs of values indicated on the MULTI-DISPLAY (LED).. 
Pairs of values:
tt X: Circulator temperature 1 or 2 or 3 (actual value TT )
The actual temperature of the bath is simultaneously saved with 
the “calibration value“ >CALVAL< and can be indicated for control 
purposes (value does not flash). 
 
Ct X: Calibration temperature 1 or 2 or 3 (actual value CT ) 
The „calibration value“ is determined with a temperature measuring 
device and saved under menu item >CALVAL<. 
(value flashes
)