JULABO MB Heating Immersion Circulator User Manual
Page 27

MB
27
8.3.
Over and Sub temperature warning function
Over temperature 
 
t High
205 °C
Sub temperature 
 
t Low
-55 °C
If the observance of a working temperature value >t 3< has to be 
supervised for a sensitive temperature application, then set over 
and sub temperature warning values. In the example below the 
setpoint >t 3< 85 °C is surrounded by the values >t High< 87 °C 
and >t Low< 83 °C. The electronics immediately register if the 
actual temperature breaches one of the set limit values. The 
resulting reaction is defined in the menu item >Aty (ALARM-
TYPE)< refer to (page 26). 
Factory settings:
t High
205 °C
t Low
-99.99 °C
1. Press the key
.
2. By pressing the
or
key select the menu > tHigh <
or > tLow <.
3. Press the
key. The integer digits flash.
4. Change the values to 87. °C and/or 83. °C by pressing the
and
key and confirm with the
key.
The decimal digits flash and can be adjusted if desired. 
Confirm once more by pressing the 
key.
See above examples.
The warning functions are only activated if the actual bath 
temperature remains within the set limit values for 3 seconds 
after switch-on. 
 
Recommendation:
Set the over temperature warning value > t High < 5 °C to 10 °C 
above the working temperature setpoint. 
Set the sub temperature warning value > t Low < 5 °C to 10 °C 
below the working temperature setpoint. 