JULABO FLW4006 Recirculating Coolers User Manual
Page 24
24
7.5.
Setting the feed pressure
max.
Set the max. permissible feed pressure (example: 2 bar) by slowly
closing the adjusting valve (13) on the rear and looking at the
manometer (5).
The max. pressure is determined by the weakest element in the
circuit (e. g. glass equipment).
7.6.
Setting the temperatures
Factory setting:
25 °C
Setting can be carried out in the start/stop condition.
1.
Press one of the keys
for a short moment.
The setpoint value instead of the actual value is indicated on the
display for about 8 seconds.
The value can now be changed.
2.
Change value:
Press
to set a higher value.
Press
to set a lower value.
Keep the keys depressed for the value to change fast.
3.
Press enter
to store the value.
7.7.
AUTOSTART ON / OFF
The recirculating cooler has been configured and supplied by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode, that the unit must enter a safe operating state after a power
failure (non-automatic start mode). This safe operating state is
indicated by „OFF“ on the LED temperature display. A complete
shutdown of the main functional elements such as compressor and
circulating pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the
start of the circulator directly by pressing the mains power switch.
Keep depressed enter
and
turn on the unit with the mains power switch.
For a short while the LED DISPLAY indicates the effective start mode:
AUTOSTART on.
AUTOSTART off.
Warning:
For supervised or unsupervised operation with the AUTOSTART function, avoid any
hazardous situation to persons or property.
The circulator does no longer conform to N.A.M.U.R. recommendations.