JULABO FLW4006 Recirculating Coolers User Manual
Page 12
![background image](/manuals/563187/12/background.png)
12
Recirculating Cooler
Cooling
FL4006 FLW4006
air cooled
water cooled
Working temperature range
°C
-20 ... +40
-20 ... +40
Temperature stability
°C
±0.5
Temperature selection:
digital
via key pad
indication
on
LED-DISPLAY
remote control via personal computer
indication on monitor
Temperature indication:
LED-DISPLAY
Resolution
°C 0.1
Temperature control
PID
1
Temperature sensor
Pt 100
Excess temperature protection
85 °C - fixed value
Low liquid level protection
float switch
Cooling capacity
Medium: Mixture water-glycol
°C
kW
+20 0 -10
+20 0 -10
4.0 1.9 0.9
4.0 1.7 0.7
Cooling compressor
1- stage
Refrigerant
R404A
Electrical connections:
Computer
interface
RS232
Alarm
output
for external alarm signal
Circulating pump:
discharge, max.
at 0 bar
l/min
60
60
pressure, max.
at 0 liter
bar
0.5 ... 6.0
0.5 ... 6.0
pressure, adjustable at 0 Liter
Feed pressure indication
bar
Manometer Manometer
Filling level indicator
sight glass
Filling volume
from ... to
liters
24 ... 30
Dimensions (WxLxH)
cm
60x76x115
Weight
kg
157
160
Ambient temperature range
°C
5 ... 40
Return flow temperature
°C
80
max.
Cooling water
Flow rate at 20 °C inlet temperature
l/min
-----
6.6
IP class according to IEC 60 529
IP
21
Mains power connection
400 V/3PNPE/50 Hz
V/ Hz
360-440/3PNPE/50
360-440/3PNPE/50
Current draw at 230 V
A
12
13
Mains power connection
230 V/3PPE/60 Hz
V/ Hz
207-253/3PPE/60 207-253/3PPE/60
Current draw at 208V / 220 V
A
16
------
All measurements have been carried out at: rated voltage and frequency,
ambient temperature: 20 °C