beautypg.com

JULABO FLW4006 Recirculating Coolers User Manual

Page 13

background image

13

Warning functions and safety installations
Excess temperature protection

85 °C - fixed value

Low liquid level protection

float switch

Alarm message

optical + audible (permanent)

Excess temperature - Warning function

75 °C

Overload protection

for compressor and pump motor

Classification according to DIN 12876-1

class I








Environmental conditions according to IEC 61 010-1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C

Air

humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuations of ±10 % are permissible.

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not for use in explosive atmosphere


Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).