JULABO FLW11006 Recirculating Coolers User Manual
Page 27

27
11. Remote
control
11.1. Setup for remote control
9
6
5
1
RS232
Check the interface parameters for both interfaces (on recirculating cooler and 
PC) and make sure they match. 
 
Interface parameters are pre-determined. 
Type RS232
Baudrate 4800
bauds
Parity even 
Handshake hardware 
handshake
 
 
11.2. Communication with a PC or a superordinated data system
 
 
If the recirculating cooler is put into remote control mode the MULTI-DISPLAY 
(LED) will read „R -OFF-„ = REMOTE STOP. The recirculating cooler is now 
operated via the computer. 
In general, the computer (master) sends commands to the recirculating cooler 
(slave). The recirculating cooler sends data (including error messages) only 
when the computer sends a query. 
 
In remote control mode: 
After a power interruption the order to start and all values which have to be 
adjusted must be resent from the personal computer via the interface. 
AUTOSTART is not possible. 
 
A transfer sequence consists of: 
 command 
out/in
command
 space
(
; Hex: 20) out/in command
 parameter
(the character separating decimals in a group is the
period) out
command
 end of file
(
; Hex: 0D) out/in command
 
 The response (data string) after an in command is always followed by a 
line feed (LF, Hex: 0A).
Important times for a command transmission:
To ensure a safe data transfer, the time gap between two commands should 
be at least 250 ms. 
The recirculating cooler automatically responds to an in command with a data 
string followed by a LF (Line Feed). The next command should only be sent 
after 10 ms. 
The commands are divided into in or out commands.
in commands:
asking for parameters to be displayed
out commands: setting
parameters
