beautypg.com

Hydro-Logic PS-200 PANEL MOUNT DUAL TDS CONTROLLER/MONITOR User Manual

Page 3

background image

6

3

INSTALLATION INSTRUCTIONS

USAGE INSTRUCTIONS

1. Remove the contents from the box.
2. Insert the controller into the panel. A square hole must be cut into the panel with dimensions
of 2-11/16 in. x 2-11/16 in. (68 mm x 68 mm). Page 8 includes a cut-out diagram.
3. Mount the controller to the panel by inserting the mounting brackets into the grooves on the
bottom and top of the controller. Fasten the brackets with the included screws.
4. View the contact diagram on the side of the controller (also on page 8).
5. Do NOT connect to a power source yet! Connect the black power cord to the contacts
#8 & #9 (110 V - 230V). It does not matter which color wire is connected to the
contacts. Screw in tight with a Phillips head screwdriver. (Note – If in the U.S. (or a country
that uses Type A or B plugs/sockets), connect the included plug adapter to the power cord).
6. If using a pump, valve, etc., connect a relay wire (not included) to contacts #13 and #14 for a
normally open (N.O.) position OR to contacts #12 and #13 for a normally closed (N.C.) position.
7. Connect a gray sensor cable by attaching the white wire to contact #1, the red to #2, and the
black to #3.
8. Connect the second sensor cable by attaching the white wire to contact #5, the red to #6, and
the black to #7. The cables are identical and can be swapped.
9. Align the pins of each sensor to each cable and attach. Screw the tightening ring closed. The
sensors are identical and can be swapped. Note which sensor is now the IN sensor and which is
the OUT sensor.
10. Insert each sensor electrode into a female ½” NPTF threaded fitting.
11. Attach the fitting connected to the OUT sensor to the product water.
12. Attach the fitting connected to the IN sensor to the feed water.
13. Ensure the orientation of the sensor is perpendicular to the flow of the water. Water
should flow equally over both pins. (If using a T, if you look through the top of the T, you
should see both pins equally side-by-side. If using the sensor in a tank, ensure that there are
no air bubbles trapped between the pins or in a fitting.)
14. To DISCONNECT the audible alarm, disconnect the blue wire from contacts #10 and #11.
15. Plug the power cord into an electrical outlet. The controller does not have a power switch
and will automatically power on when the power is connected.

WARRANTY

ONE YEAR LIMITED WARRANTY

The PS-200, including the controller and both sensors, is warranted by HM Digital, Inc.
("the Company") to the purchaser against defective materials and workmanship for
one (1) year from the date of purchase.

What is covered:

Repair parts and labor, or replacement at the Company's option.

Transportation charges for repaired of new product to be returned to the purchaser.

What is not covered:

Transportation charges for the defective product to be sent to

the Company. Any consequential damages, incidental damages, or incidental
expenses, including damages to property. This includes damages from abuse or
improper maintenance such as tampering, wear and tear, water damage, or any other
physical damage. The Company's products are not waterproof and should not be fully
submerged in water. Products with any evidence of such damage will not be repaired
or replaced. See additional note below.

How to obtain warranty performance:

Attach to the product your name, address,

description of problem, phone number, and proof of date of purchase, package and
return to:

HM Digital, Inc.

ATTN: Returns

5819 Uplander Way

Culver City, CA 90230

U.S.A.

Implied Warranties:

Any implied warranties, including implied warranties of

merchantability and fitness for a particular purpose, are limited in duration to five
years from date of purchase. Some states do not allow limitations on how long an
implied warranty lasts, so the above limitation may not apply to you. To the extent any
provision of this warranty is prohibited by federal and state law and cannot be
preempted, it shall not be applicable. This warranty gives you specific legal rights, and
you may also have other rights, which vary from state to state.

NOTE:

Warranties are product-specific. Third-party products and products deemed by

HM Digital as "accessories" are not covered under warranty. Third-party products and
accessories include, but are not limited to, batteries, fuses, mounting brackets, plug
adaptors and fittings.

IMPORTANT:

Double-check your contacts prior to connecting the controller to a

power source. Incorrect connections could result in shorting out the unit.

1. The controller will turn on when the power cord is plugged into an electrical outlet.
2. Open the cover on the front of the controller by gently pulling down.
3. There is only one display for both the IN mode (feed water) and OUT mode (product water).
The OUT mode is the default. To view the TDS level of the feed water, press the “IN MODE”
button. The display will switch to the feed water. NOTE: The display will automatically switch
back to the product water TDS after five seconds. To view the feed water display for longer,
press and hold the “IN MODE” button.
4. To turn off the controller, unplug it from the electrical outlet.