beautypg.com

HYDAC HDA 5500 User Manual

Page 14

background image

13

display

Setting

Setting
range

Pre-
setting

device

Lower display range (Low Range) Sensor 2

Lower display range limit

-999..9899

0.0

HDA 5500 - 1 - X - XX - XXX

Upper display range (High Range) Sensor 2

Upper display limit

-899..9999

10.0

HDA 5500 - 1 - X - XX - XXX

LED – light indicator Sensor 2

ON

OFF

On
Off

On

HDA 5500 - 1 - X - XX - XXX

decimal place for correction factor

(only availible with frequency/counter function)

0..0.000

0.0

HDA 5500 - 2 - X - XX - XXX

correction factor (frequency/counter function)

Flow measurements:

Settings:
f = (Fakt.) * Qmax / 60 (Hz)
Qmax : maximum flow of turbine in l/min
Fakt. : correktion factor of the turbine (Impulse/l).
decimal place to set in dEcF.
Display:
Flow = (f x 60) / FacT in l/min

Frequency measurements:
To display the frequency in Hz, follow settings must
be done:
dEc.F = 0 and FAcT. = 60

Counter function:
Settings :
decF = 0 und FacT = 1

0..9999

0.0

HDA 5500 - 2 - X - XX - XXX

Calibration of sensor zero point (Calibrate)

The actual pressure is saved as the new zero
point.This is possible in the range +/-2,5% of the
unit’s nominal pressure.

appears in the display when a calibration is carried
out in the permissible range, otherwise

is displayed.

This function is useful, for example, if there is always
a residual pressure in the system which should be
displayed as 0 bar.

Warning:

Following a zero point adjustment, for example, on a
600 bar unit, a pressure of up to 15 bar will be
displayed as 0 bar. Before any work is carried out on
the hydraulic system, ensure that the system is de-
pressurised.

YES/ NO

NO

For all devices