Disposal, Troubleshooting, Specifications – Heidolph Heating bath HB 4000 User Manual
Page 18

18
E
DISPOSAL
For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).
TROUBLESHOOTING
No heat:
a) check power cord
b) power switch (1b) broken (call service engineer)
c) max. temperature limiter dropped (refer to Chapter 31.1)
d) hot bath heater broken (call service engineer)
e) Error code Er3 displayed: you are heating your hot bath dry ! Disconnect push button (2a),
check whether temperature limiter dropped, reset if required, fill heating bath with fluid.
f) Error code Er1 and Er2 displayed: heating bath inspection by a service engineer authorized by
Heidolph is mandatory.
SPECIFICATIONS
heating power
1300 W
temperature range, hot bath
20-180°C
accuracy, hot bath
±1 K
bath temperature display
digital
heating bath temperature control
electronic
material, hot bath
V4A (1.4404)
heating bath diameter
240 mm
power supply
230V/50/60Hz oder 115V/50/60Hz
dissipated power
1300 W
weight, glassware not included
3,7 kg
size (wxdxh),
270 x 270 x 175 mm
power cord
2 m
protective class
IP 20
ambient conditions
0-40°C
at 80% rel. humidity