beautypg.com

Warning – Gast 0211 Series Lubricated Laboratory Vacuum Pumps and Compressors User Manual

Page 5

background image

5

Disposal

(Please note current regulations)

Parts of the rotary vane pumps and compressors,

shafts, iron or aluminum castings, plastic or glass

parts or bearings, may be recycled as scrap

materials.

PUMP DISASSEMBLY

1. Remove the End Plate bolts.

DO NOT at any time remove the rotor or loosen
any of the electric motor thru-bolts.

2. Remove End Plate and check for scoring on End

Plate, Rotor, and Body. Surfaces should be

smooth. If severe scoring is visible contact an

Authorized Service Facility.

3. Remove vanes.

4. Clean all surfaces with Gast recommended

Flushing solvent and dry well.

PUMP RE-ASSEMBLY

5. Apply a light coat of Gast recommended Oil

part# AD220,(an equivalent 10 wt. high detergent

oil can also be used) to the vanes.

6. Re-install vanes, noting the proper direction of

the beveled edge (Refer to exploded view).

7. Install End Plate and tighten End Plate Bolts.

8. Install new cover gasket on Filter/Muffler Assem-

bly and Lubricator.

9. Before putting the pump into service, ensure

that any external accessories have not been

damaged.

If the pump fails to produce proper vacuum or pressure,

or is excessively noisy, turn off and return unit to an

Authorized Service Facility.

Gast will NOT guarantee field-rebuilt product

performance. For performance guarantee, the product

must be returned to a Gast Authorized Service Facility.
Service Kit contents vary. Most contain vanes, gaskets

oiler wick and filter elements.

SERVICE KIT INSTALLATION

Do Not attempt to remove the rotor. It is held in place

by Loctite and can only be serviced by a Gast

Authorized Service Facility.

Disconnect electrical power supply cord before
installing Service Kit.

If product is hard wired into system, disconnect
electrical power at the circuit breaker or fuse box
before installing Service Kit.

Disconnect air supply and vent all air lines to
release pressure or vacuum.

Failure to follow these instructions can result in
death, fire or electrical shock.

WARNING

Electrical Shock Hazard