Exhaust Port
Notice
Review oven weight to ensure the foundation is adequate to hold
the weight.
4.2.1.1. Single Oven Placement Requirements
Place oven on a solid foundation such as a bench top, an optional cabinet base, or directly on
the floor.
Allow at least two inches clearance at the rear of the oven to provide proper ventilation. The
oven may be placed next to another cabinet or next to another oven, with three-quarters of an
inch clearance. The doors will still open.
Plumb and level the oven to assure proper heat distribution and operation of all mechanical
components.
Do not expose oven to excessive vibration and affix all electrical cabinets.
If placing the oven in an area where excessive particulate matter exists (such as a construction
site or coal processing center), periodically clean all its electrical compartments. Keep the
power supply within the specifications provided by Despatch, and use a line conditioner for a
facility with an unstable power supply.
4.2.1.2. Multiple Oven Placement Requirement
Stack up to two ovens vertically, with or without the optional framework supplied by
Despatch.
Use the holes in the rear oven feet to bolt the ovens together by removing the hole plugs in
the top of the oven beneath.
4.2.2.
Exhaust Connections
The RFD1-42-2E oven exhaust port is located on the rear
right of the oven (Figure 6). Install an exhaust stack from
the port to the outside of the building. Table 1 lists the
requirements for the exhaust stack for the RFD1-42-2E
oven.
Warning!
Do not use the oven in wet, corrosive or explosive atmospheres
unless this oven is specifically designed for a special
atmosphere.
Figure 6. Exhaust Port on the Rear
Right of the RFD1-42-2E Oven.