Maintain Rate


The chart below shows typical flow rate required to maintain the oxygen content at
different percentages of concentration in the RBC1-50 Burn-In oven with

nitrogen

atmosphere option installed. For example, it requires 36SCFH to keep the

oxygen

concentration down to 6½ percent in the closed oven

chamber. The

preliminary

settings can be determined from this graph. If the oxygen concentration must be
determined more accurately, use an oxygen monitor.

Oxygen Stabilization vs. Flow of N2

0

2

4

6

8

10

12

14

16

18

20

22

0

9

18

27

36

45

54

63

SCFH-N2

%O2