Damper full closed position, Damper in full open position, Other damper settings – Despatch RAD1-42 -2E with Protocol 3 User Manual
Page 14: Warning

T
HEORY OF
O
PERATION
RFD1-42-2E Oven
Owner’s Manual
14
Version 1
Copyright © 2012 by Despatch Industries.
All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.
conditions, load distribution, heat-up rates, cool-down rates, desired temperature uniformity and,
most importantly, the desired operating temperature. Carefully consider existing engineering
tradeoffs while using guidelines to determine the fresh air damper setting.
In general, the damper should be set so that the amount of fresh air flowing into the chamber
agrees with the desired operating temperature conditions. The following paragraphs show the
considerations involved with various damper position settings.
3.2.1.1. Damper Full Closed Position
The chamber achieves maximum attainable heat-up rates when the fresh air damper lies in the full
closed position. With the damper in the full closed position, the chamber will operate at the
desired temperature using the minimum amount of power. In most cases, the oven also efficiently
operates at the chamber’s maximum operating temperature when in the full closed position.
Warning!
This oven is not designed for use with flammable material. If
your process involves flammable material, see the RFD series.
3.2.1.2. Damper in Full Open Position
The chamber operates at its minimum operating temperature with the fresh air damper in full
open position.
Friction heat from the air recirculation system builds up in the chamber. This causes chamber
temperature to rise slightly even without the heating system on. The chamber reaches thermal
equilibrium temperature after the recirculation motor runs for an extended period of time.
The chamber cannot readily dissipate heat generated by friction without a fully open fresh air
damper. With the fresh air damper fully open, the thermal equilibrium temperature is the mini-
mum operating temperature of the chamber.
When the damper is in full open position, the oven may not be
able to heat to the maximum oven operating temperature.
3.2.1.3. Other Damper Settings
Damper settings can take distinct operating positions other than full closed or full open. In most
cases, uniformity and cool-down rates influence damper settings.
Chamber Uniformity
The system’s inside chamber pressure influences chamber uniformity. Pressure inside the
chamber depends on the amount of fresh air flowing into the chamber. When a large volume of
fresh air flows into the chamber, the chamber pressurizes slightly and overall temperature